Proof of Concept of an Endoscopic Sutureless Valve Sizer

Auteur : Vola, Marco; Maureira, Juan Pablo; Ruggieri, Vito Giovanni; Fuzellier, Jean-François; Campisi, Salvatore; Favre, Jean-Pierre; Gerbay, Antoine; Folliguet, Thierry A. Less
Lien vers la source : http://journals.lww.com/innovjournal/Abstract/2016/09000/Proof_of_Concept_of_an_Endoscopic_Sutureless_Valve.5.aspx

Abstract :

Objective: In this paper, we present an endoscopic expandable sizer conceived to allow thoracoscopic aortic valve replacement with a sutureless prosthesis using a dynamic sizing of the aortic annulus.

Methods: Ten aortic torsos were prepared using a five-trocar thoracoscopic setting. Once the aortotomy was performed and the aortic valve leaflets removed, the technical feasibility of the endoscopic sizing (introduction into the trocar, expansion into the aortic annulus, determination of the valve size, and retraction) with the device was assessed. In case of successful thoracoscopic sizing, endoscopic implantation of a sutureless valve (five LivaNova Perceval prosthesis and five Medtronic 3f Enable bioprosthesis) was performed. Before ascending aorta closure, we assessed the appropriate sealing of the bioprosthesis in the native annulus with camera visualization and a nerve hook inspection.

Results: All the 10 endoscopic sizings were technically feasible. The scheduled aortic sutureless valve implantations were successfully performed. In all cases, fitting and placement of the sutureless bioprosthesis in the flaccid heart was satisfactory, with no paraprosthetic leakage detectable by the nerve hook.